▼Product Introduction
DFS 0416-AT pressure electronic scanning valve is developed for small-range, high-precision application scenarios. It is a 16-channel fully independent high-performance pressure acquisition module for multiple measurements of dry, non-corrosive gases. The scanning valve integrates 16 silicon piezoresistive pressure sensors and gas path switching valves, has a common reference end, and the reference end can be externally connected. It has functions such as pressure testing, pressure calibration, and pipeline purging. Its interface and control commands are fully compatible with PSI 9116, and can be mixed with PSI 9116. It supports Labview software development functions.
Each channel of the DFS 0416-AT pressure electronic scanning valve uses a separate gas pressure sensor to ensure measurement accuracy through internal zero calibration, full-scale calibration or custom multi-point calibration and online calibration functions, and supports self-test and factory recovery.
Accessories include: power adapter (220V AC input, 24V DC output), transmission cable (customizable length) and dedicated data switch (power supply and signal transmission to the computer via one network cable, supporting up to 12 sets of simultaneous connections). Users can flexibly configure the matching host computer interface according to specific usage scenarios.
▼Performance indicators
◆ Number of channels: 16 independent channels (common reference terminal, reference terminal can be externally connected).
◆ Pressure range: ±5KPa, ±15KPa, ±30KPa, ±5PSI, ±50KPa (can be customized according to user needs)
◆ Pressure type: Differential pressure
◆ Measurement accuracy: ≤ ±0.05%FS
◆ Sampling rate: 0~500Hz/Ch (MAX, adjustable, when all channels of a single module are sampled simultaneously)
◆ Total thermal error: ±0.0008%FS/℃
◆ Overpressure capability: The test end and reference end of the product can withstand a measurement pressure twice or higher than its own range. ±5PSI can withstand a maximum overload of 10kPa ~ 300kPa (absolute pressure) without causing damage, and the influence of the reference end pressure on the accuracy is ≤±0.001%FS/PSI.
◆ Operating temperature (temperature compensation range): 0℃ ~60℃
◆ Measurement calibration: full-scale calibration, multi-point calibration, zero calibration, custom calibration, online calibration, measurement mode switching
◆ Measurement stability: ≤±0.045%FS
◆ Measurement nonlinearity: ≤ 0.015‰FS
◆ Communication interface: Support Ethernet 10/100BaseT, TCP/IP or UDP protocol, module IP can be modified
◆ Data output: 0~128 times data average adjustable; digital/analog output adjustable
◆ Trigger mode: support command trigger, external trigger (TTL level) timing trigger, support PTP time synchronization protocol, data with timestamp; support IEEE 1588 synchronization protocol, IEEE 1588 synchronization accuracy ≤60ns
◆ Filtering: Support 32-times, 64-times and other mean filtering; support high-speed data flow
◆ Power supply: 12~36V DC
◆ Working environment pressure: local atmospheric pressure
◆ Electrical interface: Aviation plug (plug model can be customized)
◆ Dimensions: 212mm* 89* mm*84mm (L*W*H)
◆ Housing: IP66 waterproof, shock and vibration resistant, noise resistant
◆ Weight: ≤2200g
◆ Gas line interface: quick-release, anti-drop type air guide needle, specifications support φ1.6mm×0.3mm (outer diameter×wall thickness), with anti-drop clamp ring (specifications can be customized)
▼Typical Applications